Planet 5G-10 Emission Analyzer

Specifications

Measurement:
NDIR (Non-Dispersive Infrared) for HC, CO, CO2
Electrochemical for NO, O2

Program Accuracy Adherance:
EPA ASM
OIML Class 0 and OIML Class 1

Gases Measured:
HC (Hydrocarbon)
CO (Carbon Monoxide)
CO2 (Carbon Dioxide)
O2 (Oxygen)
NO (Nitric Oxide)

Displayed Data:
HC
0-30,000ppm
CO
0-15%
CO2
0-20%
NO
0-5000ppm
O2
0-25%
Lambda
0.200-1.800%
A/F
0-30:1
RPM
0-15,000rpm
Temperature
0-302F (0-150C)

Size:
10" x 7.5" x 3.5" (26cm x 19cm x 9cm)

Weight: (excludes peripherals)
7lbs (3.1kg)

Power Supply:
Internal gel cell battery (rechargeable)
External 12 VDC supply (vehicle battery
or cigarette lighter)
120 VAC (240VAC) wall adapter

Battery Operating Time:
Up to 6 1/2 hours continuous

Battery Charge Time:
10 hours (approximate)

Operating Environment:
32 to 122 degrees Fahrenheit (0 to 50 Celsius
Up to 95% humidity
-1,000 to 8,000 ft (-300 to 2500 meters)

Measurement Resolution:
HC:
1ppm
CO:
0.01%
CO2:
0.01%
NO:
1ppm
O2:
0.01%
Lambda
0.001%
A/F:
0.01
RPM:
5rpm
Temperature
0.1 degree F (0.1degree C)

Measurement Accuracy:
HC:
0 to 2,000ppm
+/- 4ppm abs or +/- 3% rel
2,001 to 15,000
+/- 5% rel
15,001 to 30,000
+/- 8% rel
CO:
0.00 to 10.00%
+/- .02 abs / +/- 3% rel
10.01 to 15.00
+/- 5% rel
CO2:
0.00 to 16.00%
+/- .3 abs / +/- 3% rel
16.01 to 20.00
+/- 5% rel
O2:
0.00 to 25.00%
+/- 0.1 abs / +/- 3% rel
NO:
0 to 4,000ppm
+/- 25ppm abs or +/- 3% rel
4,001 to 5,000
+/- 5% rel

Sample Acquisition Time:
Approximately 6 seconds

Warm Up Time:
40 seconds

Memory:
Stores up to 250 individual records with ID
Records up to 1/2 hour of live data

PC Communication:
RS232 port
5G-10 can be operated from PC
Records can be downloaded to PC

Optional Printer:
Operates by AC adapter or optional battery

Specifications are nomimal and can vary between units.
Specifications are subject to change without notice.

Download Specifications